VMP-3e Multichannel Potentiostat.
Power, performance and precision all rolled into one robust multichannel potentiostat galvanostat.
16 channel electrochemical workstation with 1 A (extendable to 800 A) current capability as well as EIS Quality Indicators. This enhanced multichannel potentiostat is perfect for battery research & testing
A research-grade multichannel potentiostat / galvanostat, the new VMP-3e’s modular design, makes it perfect for wide-ranging applications. However, the ability to connect each potentiostat to an external high-current booster channel makes it especially suitable for battery research / testing.
A standard voltage range of ±10 V is extendable to -20 V to +20 V. And the standard current of ± 1 A can be brought down to 20nA – via modules that can be easily added in-situ, by the user, enabling this multichannel potentiostat / galvanostats to grow with your needs.
For high-end EIS measurements, a 1 MHz specification is complemented by BioLogic Quality Indicators, a feature previously only available on BioLogic Premium potentiostats. Quality Indicators enable scientists to quickly and efficiently validate their impedance measurements. Only BioLogic offers three Quality Indicators, making this the most comprehensive potentiostat range available.*
Additional features include CE to GND mode, unique to BioLogic, which enables users to choose from independent channel or multi-electrode configurations and LAN connectivity to improve multi-user working.
Powered by EC-Lab® , the advanced, ergonomic software that has earned itself the reputation as a global benchmark for potentiostat control software, the VMP-3e boasts a vast array of unique functionality. Only EC-Lab enables users to build and modify experiments incrementally with “modify-on-the-fly”. But a vast array of presets is available, if necessary.
The VMP-3e is a true benchmark electrochemical workstation. A multichannel potentiostat / galvanostat which, when combined with the power of BioLogic’s EC-Lab® instrument control and data analysis software, will help take your research to the next level.
*Total Harmonic Distortion (THD), Non-Stationary Distortion (NSD), Noise to Signal Ratio (NSR)
Powerful, modular and easy to use potentiostat control and analysis software
The most powerful multichannel potentiostat galvanostat is only as performant as the software that sits behind it. BioLogic’s EC-Lab® has earned itself the reputation as the benchmark for potentiostat control software based on a combination of intuitive control and analytic power.
General
General | |
---|---|
Slots available | Up to 16 |
Electrodes connections | 2, 3, 4, 5 |
Compliance | 20 V adjustable from ± 10 V between [-20 ; 20] V |
Acquisition time | 20 µs with EC-Lab developer package, 200 µs with EC-Lab. Live processed data : 200 µs, independent of sampling rate (Q charge, Q discharge, Energy, Power...) |
Software | EC-Lab, EC-Lab developer package |
Add-ons compatibility | Yes |
Dimension (W x D x H) | 448 x 455 x 280 mm |
Weight | 15 kg |
Power requirement | 1000 W |
Interfaces | Ethernet (Direct, LAN, Wifi), USB 2.0 |
Channel specifications
Voltage | ||
Ranges | Customizable ranges. Compliance of 20 V adjustable (example 0 - 20 V). Up to 60 V with FlexP-0060 | ± 2,5 V, ± 5 V, ± 10 V. ± 10 V adjustable (example 0 - 20 V). Up to 60 V with FlexP0060/FlexP0160 |
Accuracy | < 0.1 % FSR* (20 mV) | < 0.1 % FSR* (20 mV) |
Resolution | 5 µV to 305 µV | 76 µV below 10 V, 152 µV from 10 to 15 V, 305 µV from 15 to 20 V |
Current | ||
Ranges | ||
Accuracy | ||
Resolution | ||
EIS | ||
EIS capability | ||
EIS quality indicators |
*FSR = Full Scale Range
Advanced
Advanced | |
---|---|
Floating options | CE to ground |
Permeation (WE to ground) | No |
Multi-electrode (CE to ground) | Yes (with "e" type channel board) |
Filters | Software |
Input/ouput | 2x analog inputs 1x analog output 3x TTL (trigger input/output, and safety) 2x monitor outputs (current and voltage) |
Stability control mode | 7 bandwidths |
Calibration | More than 50 verified points. |